Sealing parameters

Technical data and parameters for heat sealing laboratory plates with the IntelliXseal™ SA Semi-Automatic Sheet Heat Sealer (formerly 4s3™)

Product name/code Temperature & dwell time for
96 well PCR plates 384 well PCR plates Vision Plates™ (COC plates)
Clear Heat Seal/4ti-0541 175-185 °C, 2-3 sec 165-180 °C, 3 sec 185-200 °C, 3 sec
Clear Weld Heat Seal Mark 2/4ti-0575 175-185 °C, 2-3 sec 170-175 °C, 2-3 sec -
Clear Heat Seal Plus/4ti-05481 175-185 °C, 2-3 sec 165-180 °C, 3 sec 185-200 °C, 3 sec
Clear Heat Seal 3730/4ti-0581 165-175 °C, 3 sec 165-175 °C, 2 sec 175-185 °C, 2-3 sec
Peel Heat Seal/4ti-0521 175-185 °C, 3 sec 170-175 °C, 2-3 sec -
Universal Peel Heat Seal/4ti-05231 175 °C, 2 sec 175 °C, 2 sec 180 °C, 2 sec
DMSO Resistant Peel Heat Seal/4ti-0587 175-185 °C, 3 sec 170-175 °C, 2-3 sec -
Pierce Heat Seal/4ti-0531 160-175 °C, 2 sec 160-175 °C, 2 sec 185-200 °C, 3 sec
Pierce Heat Seal Strong/4ti-05381 170-180 °C, 2 sec 170-180 °C, 2 sec 180-200 °C, 3 sec
Random Access Pierce Heat Seal Strong/4ti-05381/RA 170-180 °C, 2 sec 170-180 °C, 2 sec 180-200 °C, 3 sec
96 INDI™ Heat Seal/4ti-0543 175 °C, 3 sec - -
Foil Heat Seal/4ti-0536 165-180 °C, 2 sec 165-175 °C, 2-3 sec 185-200 °C, 3 sec
Polystyrene Foil Heat Seal/4ti-0547 165-180 °C, 2 sec 165-175 °C, 2-3 sec 185-200 °C, 3 sec
Thermal Bond Heat Seal/4ti-0591 170-180 °C, 2-3 sec 160-170 °C, 2 sec -
Gas Permeable Heat Seal Mark 2/4ti-0597 170 °C, 2 sec 170 °C, 2 sec 170 °C, 2 sec
Gas Permeable Clear Heat Seal/4ti-0541/SLIT 175-185 °C, 2-3 sec 165-180 °C, 3 sec 185-200 °C, 3 sec